KB0013: BioLab 3.0.x Error -200361

Issue

There is a compatibility issue when running Windows 7, or newer, with a BioNex 2-Slot and BioLab 3.0.x. This results in a form of the error:

The following error has occurred:

Error -200361 occurred at DAQmx Read (Analog 1D DBL NChan 1Samp).vi:3

Possible reason(s):

Onboard device memory overflow. Because of system and/or bus-bandwidth limitations, the driver could not read data from the device fast enough to keep up with the device throughput.

Reduce the sample rate, or reduce the number of programs your computer is executing concurrently.

Task Name: _unnamedTask<FA>

Please check your setup and try again.

Solution

Install the DAQmx 9.2.3 Patch available below.